❏ 7-inch color touch control interface display
▸ 7-inch color touch control interface display, an interface can be real-time display of inflow and downflow air velocity, fan operation time schedule, the status of the front window, the filter and the sterilization lamp life percentage, the working environment temperature, the output and shutdown operation of the socket, lighting, sterilization and fan, the operation log and alarm function, without the need to switch the interface
❏ Energy-efficient DC brushless constant airflow fan
▸ Energy-efficient design with ultra-low-energy DC motor saves 70% energy consumption (compared to traditional AC motor designs) and reduces heat emissions
▸ Real-time airflow regulation ensures that inflow and outflow velocities remain stable, with air velocity sensors monitoring airflow measurements through the work zone. The airflow can be adjusted to compensate for changes in filter resistance
▸ No need to turn off the machine when the experimental process needs to be paused, closing the front window automatically enters a low-speed energy-saving mode of operation, the safety cabinet can be operated in 30% energy-saving mode to maintain the cleanliness of the operating area, reduce the power consumption of the operation and the energy-saving mode of the percentage of the adjustable. Once the front window is opened, the cabinet enters into normal operation, effectively improving operational efficiency
▸ With a power failure memory protection function, such as accidental power failure, power can be restored to return to the operating state before the power failure, fully protect personnel safety
❏ Humanized structure design
▸ Front-end 10 ° tilt design, more in line with ergonomics, so that the operator is comfortable and not oppressed
▸ Extra-large color touch screen display, providing English language interface, one click to turn off the alarm beeping function
▸ The whole piece of worktop and sidewall are made of 304 stainless steel, safe, reliable and easy to clean.
▸ Concealed lighting, avoiding personnel to look directly at the light source from the front of the eyes, to reduce the harm to eyesight
▸ Tool-less removal/installation of the work surface, easy to clean the liquid collection tank
▸ Brakeable mobile casters provide convenience for moving the position and at the same time provide security for the fixed installation position
❏ High-quality ULPA filter
▸ ULPA filters with high-efficiency, low-pressure-drop, high-strength, and low-boron air cartridges reduce pressure drop while extending filter life, and the filtration efficiency can reach 99.9995% for particle sizes up to 0.12μm
▸ Both supply and exhaust filters are equipped with unique “Leakage Stop” technology, which ensures that the air is clean to ISO Class 4
❏ Sterilization by appointment
▸ Users can directly turn on the UV sterilization, you can also make an appointment for sterilization, set up the sterilization appointment time, the biological safety cabinet will automatically enter the sterilization appointment state, with the ability to set up the appointment for Monday to Sunday, the start and end time of the sterilization function
▸ UV lamp and front window interlock function, only after closing the front window, you can open the UV sterilization, in the sterilization process, when the front window is opened, the sterilization is automatically closed to protect the experimenter or sample
▸ UV lamp and lighting interlock function, when the UV lamp is turned on, the lighting is automatically turned off
▸ With power failure memory protection, when the power failure recovery, the safety cabinet can quickly enter the sterilization state
❏ Three levels of authority user management function
▸ Three levels of authority users include administrators, testers and operators, corresponding to different use of the operating privileges, only the administrator has all the use of the operating privileges for the safe management of the laboratory to provide the convenience of the laboratory, can provide more than five user roles
❏ Logging function
▸ Log records include operation logs, alarm logs, historical data and historical curves, and you can view the last 4,000 operation logs and alarm logs, the last 10,000 historical data, as well as the historical operating curves of the inflow and down flow velocity
▸ The administrator can manually delete the operation log, alarm log, and historical data
▸ When the fan is turned on, the historical data is sampled according to the set sampling interval, which can be set between 20 to 6000 seconds.