AS1500A2 Biosafety Cabinet
❏ 7-inch touchscreen HMI interface
▸ 7-inch touch screen human-machine interactive interface, one interface can real-time display the time progress of the fan operation, the status of the front window, the percentage of the life of the filter and the sterilization lamp, control the output and shutdown operation of the socket, the lighting, the sterilization, and the fan, view the operation log and provide an alarm function without switching the interface
❏ Energy-efficient DC brushless constant airflow fan
▸ Energy-efficient design - advanced design with ultra-low energy consumption, saving 60% energy consumption (compared to traditional AC motor design); 60% reduction in heat emission (compared to traditional AC motor design); unique energy-saving mode (10% energy-saving mode for energy consumption)
▸ Real-time airflow regulation ensures that inflow and downflow velocities remain constant. An air velocity sensor filter monitors airflow measurements through the work area. Airflow can be adjusted to compensate for protection in response to changes in filter resistance
❏ Humanized structural design
▸ Moderate height, wind speed type velocimeter.
▸ Extra-large microcomputer control system panel, humanized tilting front window design
▸ Comfortable hand shelf, wide working area
▸ Whole-piece work surface, extraordinary safety
▸ Front window wash position designed to reduce maintenance frequency
❏ Unique value machine and power failure memory function
▸ No need to turn off the machine when the experimental process needs to be paused, closing the front window automatically enters the low-speed energy-saving operation state to maintain the cleanliness of the operation area. Once the front window is opened, the machine immediately enters the normal operation state, which effectively improves the operation efficiency. If there is an accidental power failure, it can return to the operation state before the power failure after restoring the power supply, which fully protects the safety of personnel.
❏ High-quality ULPA filter
▸ Adoption of high-efficiency, low-pressure-drop, high-strength and low-boron content air filter cartridges prolongs the service life of the filter while reducing the pressure drop
▸ Unique “Leak Stop” technology for both supply and exhaust air filters, ensuring ISO 4 (Class 10) air cleanliness.
❏ Sterilization function by appointment
▸ Users can turn on one-key sterilization reservation, set the sterilization reservation time, click on the sterilization reservation switch, a prompt box will pop up to enter the sterilization reservation status, click OK and the system will enter the sterilization reservation status.
AirSafe 1500 (A2) | 1 |
Power Cord | 1 |
Fuse | 2 |
Product Manual, Test Report, etc. | 1 |
Cat.No. | AS1500 |
Filtration efficiency | >99.9995%, @0.12μm |
Air supply and exhaust filters | ULPA filters |
Air Cleanliness | ISO 4 Class |
Execution standards | YY0569(GB 4793.1、GB/T 18268.1) |
Down flow velocity | 0.35m/s |
Inflow velocity | 0.55m/s |
Noise level | <65 dB |
Vibration | <5μm(center of tabletop) |
Personnel Protection | A.Total colony in impaction sampler <10CFU./time B. Total colony in slot sampler <5CFU./time |
Product Protection | Total colony in culture dish <5CFU./time |
Cross-contamination Protection | Total colony in culture dish <2CFU./time |
Max Consumption(with spare socket) | 1.65KW |
Rated Power(without spare socket) | 0.33KW |
Internal Dimensions | 1380×580×798mm |
External Dimension | 1500×810×2290mm |
Support base (with wheel) | 1485×710×670mm |
Size and Qty. of Light | 18W×1 |
Size and Qty. of UV Light | 20W×1 |
Lumin. | ≥900LX |
Cabinet Materials | High-grade steel and lacquered in ivory |
Working Area Materials | SS304 fully finished |
Air Direction | Top out |
Power supply | 115/230V±10%, 50/60Hz |
Weight | 308kg |
Cat. No. | Product name | Shipping dimensions W×D×H (mm) |
Shipping weight (kg) |
AS1500 | Biosafety Cabinet | 1710×870×1770mm | 345 |
♦ Elevating Viral Research Safety: AS1500A2 at Wuhan Institute of Virology
The AS1500A2 Biosafety Cabinet plays a critical role at the Wuhan Institute of Virology, a leading research center dedicated to virology and innovation. As the home of China's only P4 biosafety laboratory, the institute provides fundamental research for infectious disease prevention. With precise airflow control and ULPA filtration, the AS1500A2 ensures top-tier protection for personnel and samples during experimental workflows. This partnership supports groundbreaking research into high-pathogenic viruses, advancing efforts to combat global infectious disease threats with cutting-edge technology and innovation.
♦ Supporting Epigenetics Studies: AS1500A2 at Anhui Medical University
At Anhui Medical University’s Basic Medical School, the AS1500A2 Biosafety Cabinet enables groundbreaking research into piRNA regulation in germ cells, epigenetic modifications in aging, and nucleolar structure in gene expression. With stable inflow and downflow, the AS1500A2 ensures optimal protection for researchers and samples. The ULPA filtration system guarantees ultra-clean air, creating a controlled environment that enhances the accuracy of experimental outcomes. By fostering a secure and sterile setting, the AS1500A2 empowers researchers to delve deeper into the molecular mechanisms driving human health.
♦ Advancing Hepatitis C and Virus Research: AS1500A2 at Guangzhou Laboratory
The AS1500A2 Biosafety Cabinet is an essential tool at Guangzhou Laboratory, known for pioneering the first immune-competent animal model for chronic HCV infection. Their research has unveiled the molecular mechanisms linking HCV to metabolic disorders and focuses on high-pathogenic viruses, vaccine development, and innovative diagnostics. The AS1500A2’s ULPA filtration and airflow precision ensure unparalleled sample and personnel protection. By providing a secure environment, the AS1500A2 contributes to transformative research aimed at uncovering viral pathogenesis and developing new therapeutic solutions.